Ti-6Al-7Nb
Ti-6Al-7Nb contains 6% aluminum and 7% niobium, which was developed as a biomedical replacement for Ti-6Al-4V alloy because Ti-6Al-4V contains vanadium, an element that has demonstrated cytotoxic outcomes when isolated.
Ti-6Al-7Nb is a dedicated high strength titanium alloy with excellent biocompatibility for surgical implants.
Ti-6Al-7Nb alloy is double or triple vacuum melted to provide an ingot composition that is very uniform and homogeneous. The ingot is hot pressed, and finished into round and flat bar products using conventional titanium alloy processing methods.
Chemical composition of Ti-6Al-7Nb (%, max.)
Grade
UNS
O
N
C
H
Fe
Ta
Al
Nb
Ti-6Al-7Nb
R56700
0.20
0.05
0.08
0.009
0.25
0.50
6.5-7.5
5.5-6.5
Mechanical properties of Ti-6Al-7Nb
Grade
Tensile strength
Yield strength
Elongation in 4D
Reduction of area
Hardness
Ti-6Al-7Nb
900
800
10
25
32
(MPa, min.)
(MPa, min.)
(%, min.)
(%, min.)
(HRC, max.)
Available forms
● Ingot
● Forging
● Rod and bar
● Plate and sheet
● Wire
● Custom-made
Specification
ASTM F1295
ISO 5832-11
Typical application
Surgical implants such as hip joints and dental implants to replace Ti-6Al-4V